    Micro Raman Spectrometer

    Model :MR532/785/1064

    Keyword :Micro Raman, Micro Raman Spectrometer

    Technical Support:

    400-820-7939

    Technical Support:

    021-50687686

    • Introduction

    Introduction

    MR532/785/1064 Micro Raman Spectrometer is composed of Raman three-channel module, LED light source, metallographic microscope and other parts. The operation is simple, the acquisition speed is fast, and the microstructure and Raman spectrum information of the sample can be observed in a short time, which can quickly help users to detect and analyze the sample. It is widely used in scientific research fields such as biomedicine, photoluminescence, microplastics, and two-dimensional material graphene.

    For example, the Raman three-channel module developed by Oceanhood can support Raman spectroscopy measurement of any three different excitation wavelengths, such as 532nm, 633nm, 785nm, 830nm, and 1064nm. The module is equipped with an optical path switching system, which can perform rapid switching of excitation at different wavelengths through software, and can superimpose and compare the collected Raman spectra. The probe on the module can be plug-and-play, which is convenient for the detection of unconventional samples. The Raman three-channel module can be customized and adapted according to the user's existing microscope, and assembled into a micro-Raman spectrometer.

    The micro-Raman spectrometer can support the expansion of the Mapping platform (that is, the Raman spectral imaging function), and can perform Raman spectral imaging on the sample micro area. Visually display the chemical composition distribution, surface physicochemical properties and more information of the sample by means of images. The high-precision mobiles platform, with a minimum step of 1μm, can accurately detect the Raman signal of the sample. The software has perfect functions, which can realize auto focus function, bright field image stitching and mapping image display.

    Feature

    ◆ Three-channel excitation: fast switching of various excitation wavelengths;

    ◆ fast detection speed: can quickly obtain the microstructure and Raman spectrum information of the sample;

    ◆ full software functions: autofocus, spectrum stacking, brightfield image Stitching and mapping image display;

    ◆ Excellent spectral performance: the instrument has excellent reproducibility, sensitivity, resolution and other performance;

    ◆ Support custom Raman three-channel module, adapt to existing micro

    Description

    MR532Product parameter
    Item Value
    Spectral range 200-3200cm-1
    Resolution(FWHM) <8cm-1 @25μm Slit
    Spectral shift error <3cm-1
    Spectral frequency shift repeatability <1cm-1
    Excitation wavelength 532±1nm,line width≤0.2nm
    Integral time 8ms-30min
    Life 5000hrs
    objective lens Infinity long working distance plan achromatic metallographic objective 10X 20X 50X
    converter Internal positioning 5-hole converter
    CCD imaging Imageable
    Stage Double-layer mechanical stage, size: 210mm×140mm; moving range: 63mm×50mm
    Epi-illumination system 6V30W Halogen light, adjustable brightness
    Focusing structure Coaxial coarse and fine adjustment, equipped with limit device and locking device, coarse stroke per revolution: 30mm, fine adjustment handwheel grid value 2μm
    Dimensions Raman Module: 300×210×80mm, Micro Module: 550×280×380mm
    Working temperature 0-45℃
    Working humidity 5%-80%
    MR785Product parameter
    Item Value
    Spectral range 200-3000cm-1
    Resolution(FWHM) <8cm-1 @25μm Slit
    Spectral shift error <3cm-1
    Spectral frequency shift repeatability <1cm-1
    Excitation wavelength 785±0.5nm,line width≤0.08nm
    Integral time 8ms-30min
    Life 10000hrs
    objective lens Infinity long working distance plan achromatic metallographic objective 10X 20X 50X
    converter Internal positioning 5-hole converter
    CCD imaging Imageable
    Stage Double-layer mechanical stage, size: 210mm×140mm; moving range: 63mm×50mm
    Epi-illumination system 6V30W halogen lamp, brightness adjustable
    Focusing structure Coaxial coarse and fine adjustment, equipped with limit device and locking device, coarse stroke per revolution: 30mm, fine adjustment handwheel grid value 2μm
    Dimensions Raman Module: 300×210×80mm, Micro Module: 550×280×380mm
    Working temperature 0-45℃
    Working humidity 5%-80%
    MR1064Product parameter
    Item Value
    Spectral range 200-2500cm-1
    Resolution(FWHM) <15cm-1 @25μm Slit
    Spectral shift error <3cm-1
    Spectral frequency shift repeatability <1cm-1
    Excitation wavelength 1064±0.5nm,line width≤0.2nm
    Integral time 8ms-30min
    Life 10000hrs
    Laser power stability ≤3% P-P(@2hrs)
    objective lens Infinity long working distance plan achromatic metallographic objective 10X 20X 50X
    converter Internal positioning 5-hole converter
    CCD imaging Imageable
    Stage Double-layer mechanical stage, size: 210mm×140mm; moving range: 63mm×50mm
    Epi-illumination system 6V30W halogen lamp, brightness adjustable
    Focusing structure Coaxial coarse and fine adjustment, equipped with limit device and locking device, coarse stroke per revolution: 30mm, fine adjustment handwheel grid value 2μm
    Dimensions Raman Module: 300×210×80mm, Micro Module: 550×280×380mm
    Working temperature 0-45℃
    Working humidity 5%-80%